OneTouch Verio Reflect® – Result Variations
Products
OT Verio Reflect Meter
Result Variations:
- Your OneTouch Verio Reflect® Meter result may vary slightly from your actual blood glucose value. This may be due to slight differences in technique and the natural variation in the test technology.
- The chart below shows the results of a study where 354 typical users used the OneTouch Verio Reflect® Meter to test their blood glucose level. For example, in this study, the OneTouch Verio Reflect® Meter gave results within 15% of their true blood glucose level 351 out of 354 times.
Regression Statistics:

- Samples were tested on one of three test strip lots. Results indicate that the OneTouch Verio Reflect® System compares well with a laboratory method.
Precision:
- Within Run Precision (300 Venous Blood Tests per glucose level)
-
- Data generated using the OneTouch Verio Reflect® Meter show that the greatest variability observed between test strips from three strip lots when tested with blood is 1.05 mg/dL SD or less at glucose levels less than 75 mg/dL or 3.07% CV or less at glucose levels at 75 mg/dL or above.
- Total Precision (600 Control Solution Tests per Glucose Level)
- Data generated using the OneTouch Verio Reflect® Meter.
AW 07201105A