OneTouch Verio Reflect® - Error Messages
Products
OT Verio Reflect Meter
OneTouch Verio Reflect® Error Messages:
- Error Message 1
- Error Message 2
- Error Message 3
- Error Message 4
- Error Message 5
- Error Screens
- Useful Links
- What it means:
- There is a problem with the meter.
- What to do:
- Remove the batteries.
- Press the Power or Ok button for 3 seconds.
- Reseat the batteries.
- Turn meter back on normally (with strip or Power button).
- If the problem persists, contact OneTouch® Customer Care.
- What it means:
- Error message could be caused by a used test strip, applying blood to the test strip before inserting it into the meter or a problem with the meter or test strip.
- Error 2.1 can be caused by inserting a test strip which has already been used.
- Error 2.2 can be caused by the test strips being stored incorrectly.
- What to do:
- Check test strip port for contamination. Confirm if any control solution or blood have gone inside the test strip port.
- Repeat the test with a new test strip.
- If the problem persists, contact OneTouch® Customer Care.
- What it means:
- The sample was applied before the meter was ready.
- What to do:
- Repeat the test with a new test strip. Apply a blood or control solution sample only after the Apply Blood or Apply Control Solution screen appears on the display.
- If this message continues to appear, contact OneTouch® Customer Care.
- What it means:
- One of the following may apply:
- Not enough blood or control solution was applied or more was added after the meter began to count down.
- The test strip may have been damaged or moved during testing.
- The sample was improperly applied.
- There may be a problem with the meter.
- Error 4, sub-code 2 can be caused by testing using control solution without setting the meter to display “Apply Control Solution”.
- One of the following may apply:
- What to do:
- If testing with control solution, go into the meter Settings and select Control Solution. Then insert the test strip and wait for “Apply Control Solution” to appear on the screen.
- Repeat the test with a new test strip.
- If the error message appears again, contact OneTouch® Customer Care.
- What it means:
- The meter has detected a problem with the test strip. Possible cause is test strip damage.
- What to do:
- Repeat the test with a new test strip.
- If the error message appears again, contact OneTouch® Customer Care.
- If you cannot resolve the error with your meter, contact OneTouch® Customer Care. They will refer to the error number, and a sub-error code found in the Meter Info screen, to help troubleshoot the problem.
- For more information on viewing the Meter Info screen:
- OneTouch Verio Reflect® - Preparing The Meter For A Blood Glucose Test - Notes and Cautions
- OneTouch Verio Reflect® - Preparing The Meter For A Blood Glucose Test
- OneTouch Verio Reflect® - Testing with Control Solution
- OneTouch Verio Reflect® - Error and Other Messages
- OneTouch Verio Reflect® - Applying Blood And Reading Results
- OneTouch Verio Reflect® - Getting To Know Your OneTouch® Delica® Plus Lancing Device
- OneTouch Verio Reflect® - Preparing Your OneTouch® Delica® Plus Lancing Device
- OneTouch Verio Reflect® - Sampling Blood From The Fingertip
- OneTouch Verio® Test Strips - General Information
AW 07201105AFILE-REC-0124803 03FILE-REC-0126026