OneTouch Verio Reflect® – Error 4
Products
OT Verio Reflect Meter
Error 4:
- What it means:
- One of the following may apply:
- Not enough blood or control solution was applied or more was added after the meter began to count down.
- The test strip may have been damaged or moved during testing.
- The sample was improperly applied.
- There may be a problem with the meter.
- Error 4, sub-code 2 can be caused by testing using control solution without setting the meter to display “Apply Control Solution”.
- One of the following may apply:
- What to do:
- If testing with control solution, go into the meter Settings and select Control Solution. Then insert the test strip and wait for “Apply Control Solution” to appear on the screen.
- Repeat the test with a new test strip.
- OneTouch Verio Reflect® – Preparing The Meter For A Blood Glucose Test - Notes and Cautions
- OneTouch Verio Reflect® - Preparing The Meter For A Blood Glucose Test
- OneTouch Verio Reflect® - Sampling Blood From The Fingertip
- OneTouch Verio Reflect® – Applying Blood And Reading Results
- OneTouch Verio® Test Strips - General Information
- If the error message appears again, contact OneTouch® Customer Care.
Error Screens:
- If there is a problem with the meter, error screens may appear. If you cannot resolve the error with your meter, contact OneTouch® Customer Care. They will refer to the error number, and a sub-error code found in the Meter Info screen, to help troubleshoot the problem.
- For more information on viewing the Meter info screen:
AW 07201105A FILE-REC-0126026