OneTouch Verio Flex® – Lay User Accuracy

Products OT Verio Flex Meter

Lay User Accuracy:

  • User Accuracy of the system was assessed at 3 clinical sites. 172 patients tested their blood using the OneTouch Verio Flex® System where results were compared to the YSI 2300 Glucose Analyzer laboratory Instrument.


 Fingertip Results for Glucose Concentrations <75 mg/dL:

  • Percent (and number) of meter results that match the laboratory test.

 

Fingertip Results for Glucose Concentrations ≥75 mg/dL:

  • Percent (and number) of meter results that match the laboratory test.


Regression Statistics:

  • Samples were tested on each of three test strip lots. Results indicate that the OneTouch Verio Flex® System compares well with a laboratory method.
image.png


Precision:

  • Within Run Precision (300 Venous Blood Samples Tested per Glucose Level).
    • Data generated using the OneTouch Verio Flex® Meter.
image.png
  • Results show that the greatest variability observed between test strips when tested with blood is 0.83 mg/dL SD or less at glucose levels less than 75 mg/dL or 1.98% CV or less at glucose levels at 75 mg/dL or above.
  • Total Precision (600 Control Solution Tests per Glucose Level).
    • Data generated using the OneTouch Verio Flex® Meter.
image.png

AW 06932408 B