OneTouch Ultra Plus Flex™ - Lay User Accuracy, Regression Statistics And Precision
Products
OT Ultra Plus Flex Meter
Lay User Accuracy:
Regression Statistics:
Precision:
- User Accuracy of the system was assessed at 4 clinical sites. 169 patients tested their blood using the OneTouch Ultra Plus Flex™ System where results were compared to the YSI 2300 Glucose Analyzer laboratory Instrument.
- Fingertip Results for Glucose Concentrations <75 mg/dL
- Percent (and number) of meter results that match the laboratory test
- Fingertip Results for Glucose Concentrations ≥75 mg/dL
- Percent (and number) of meter results that match the laboratory test
Regression Statistics:
- Samples were tested on each of three test strip lots. Results indicate that the OneTouch Ultra Plus Flex™ System compares well with a laboratory method.
Precision:
- Within Run Precision (300 Venous Blood Samples Tested per Glucose Level)
- Data generated using the OneTouch Ultra Plus Flex™ Meter.
- Results show that the greatest variability observed between test strips when tested with blood is 0.86 mg/dL SD or less at glucose levels less than 75 mg/dL or 2.13% CV or less at glucose levels at 75 mg/dL or above.
- Total Precision (600 Control Solution Tests per Glucose Level)
- Data generated using the OneTouch Ultra Plus Flex™ Meter.
AW 07295104 B