OneTouch VerioIQ® - System Accuracy
Products
OT Verio IQ Meter
System Accuracy:
System Accuracy Results for Glucose Concentrations >75 mg/dL:
System Accuracy Results across the entire Glucose Range:
Precision:


- Diabetes experts have suggested that glucose meters should agree within 15 mg/dL of a laboratory method when the glucose concentration is lower than 75 mg/dL, and within 20% of a laboratory method when the glucose concentration is 75 mg/dL or higher. Samples from 100 patients were tested using both the OneTouch Verio IQ® System and the YSI 2300 Glucose Analyzer laboratory instrument.
- Percent (and number) of meter results that match the laboratory test
- Percent (and number) of meter results that match the laboratory test
- Percent (and number) of meter results that match the laboratory test
- Therefore, 100% of the total results obtained with the OneTouch Verio IQ® System achieved the goal suggested by the diabetes experts.
- Samples were tested in duplicate on each of three test strip lots. Results indicate that the OneTouch Verio IQ® System compares well with a laboratory method.
- Within Run Precision (300 Venous Blood Tests)
- Results show that the greatest variability observed between test strips when tested with blood is 2.86% or less.
- Total Precision (600 Control Solution Tests)
- Fingertip Results for Glucose Concentrations <75 mg/dL
- Percent (and number) of meter results that match the laboratory test
- Fingertip Results for Glucose Concentrations >75 mg/dL
- Percent (and number) of meter results that match the laboratory test
AW 06703306A