OneTouch Verio® – System Accuracy

Products OT Verio Meter
System Accuracy:

 

  • Diabetes experts have suggested that glucose meters should agree within 15 mg/dL of a laboratory method when the glucose concentration is lower than 75 mg/dL, and within 20% of a laboratory method when the glucose concentration is ≥75 mg/dL or higher. Samples from 100 patients were tested using both the OneTouch Verio® System and the YSI 2300 Glucose Analyzer laboratory instrument.

System Accuracy Results for Glucose Concentrations <75 mg/dL:
  • Percent (and number) of meter results that match the laboratory test:
image.png

System Accuracy Results for Glucose Concentrations ≥75 mg/dL:
  • Percent (and number) of meter results that match the laboratory test:
image.png

System Accuracy Results across the entire Glucose Range:
  • Percent (and number) of meter results that match the laboratory test:
image.png
  • Therefore, 100% of the total results obtained with the OneTouch Verio® System achieved the goal suggested by the diabetes experts.

Fingertip Results for Glucose Concentrations <75 mg/dL:
  • Percent (and number) of meter results that match the laboratory test:
image.png

Fingertip Results for Glucose Concentrations ≥75 mg/dL:
  • Percent (and number) of meter results that match the laboratory test:
image.png

Regression Statistics:
  • Samples were tested in duplicate on each of three test strip lots. Results indicate that the OneTouch Verio® System compares well with a laboratory method.
image.png

Precision:
  • Within Run Precision (300 Venous Blood Tests per Glucose Level)
    • Results show that the greatest variability observed between test strips when tested with blood is 2.44% or less.
image.png
  • Total Precision (200 Control Solution Tests per Glucose Level)
image.png

AW 06908604A